0
2011 Design of Medical Devices Conference Abstracts

Design of a Novel Experimental Setup for the Assessment of the Fossa Ovalis Within Large Mammalian Hearts: Investigating Tissue Properties and Clinical Devices Used for Transseptal Access OPEN ACCESS

[+] Author and Article Information
Stephen Howard, Cory Kasprzak, Chris Rolfes, Paul A. Iaizzo

University of Minnesota

J. Med. Devices 5(2), 027539 (Jun 15, 2011) (1 page) doi:10.1115/1.3591402 History: Published June 15, 2011; Online June 15, 2011
FIGURES IN THIS ARTICLE

Clinical access to the left atrium and/or to the left ventricle for medical device delivery, from a nonsurgical standpoint, continues to be a challenge. Currently such procedures involve left heart access via delivery through the arterial system via the aorta or across the atrial septal wall. More specifically, medical devices delivered through the atrial septum require transseptal punctures by utilizing tools and delivery systems, which include puncturing needles, stiff wires, or more sophisticated approaches, such as wire tips using rf energy. Typically, from the right atrium, one hopes to make this approach through the fossa ovalis and thus gain access to the left atrium. Next, a dilator and an outer device delivery sheath/catheter (clinically available between 7°F and 24°F) are employed to pass through the initial puncture and then create a larger hole through the septum. With continued advancements in intracardiac device technologies, it is foreseeable that larger and larger tools may be needed to perform more complicated procedures from such a percutaneous approach (e.g., the transcatheter deliveries of mitral and/or aortic valves). One of the primary aims of the present study was to assess the relative properties of the fossa ovalis within a large sample of large mammalian hearts and/or the relative amount of forces needed to induce anatomical impacts (i.e., tenting, puncturing, and dilating the fossa ovalis). To do so, an experimental platform has been uniquely developed and experiment data collected. Briefly, the interatrial septums from large mammalian hearts were excised and placed on a plate lined with a silicone elastomer. This plate contains a hole where the fossa is centered and then pinned onto the gel. A smaller plate covers the tissue and is also pinned to the plate, preventing the pins from bending or leaning inward as the tissue is being tested. The prepared sample can then be depressed with a rod attached to a force transducer to predetermined distances and cause strain on the tissue: Resulting forces are digitally recorded. Such experimental protocols can be performed multiple times and to date have allowed for consistent measurements of each fossa ovalis tested. This experimental approach has also allowed us to determine the amount of force required to perforate a given fossa ovalis and thus has provided us with insights relative to device designs. Further use of this setup could be employed to study the transseptal passage and imposed forces on larger transseptal devices; e.g., to determine their potential impacts on the native anatomy of the heart. This novel approach to examine the anatomical and physical properties of the cardiac fossa ovalis should be a great value for those designing or clinically deploying transseptal therapies.

Copyright © 2011 by American Society of Mechanical Engineers
This article is only available in the PDF format.

References

Figures

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In