Sensorless Force Sensing for Minimally Invasive Surgery

[+] Author and Article Information
Baoliang Zhao

Department of Mechanical
and Materials Engineering,
University of Nebraska–Lincoln,
Lincoln, NE 68508
e-mail: baoliang.zhao2@gmail.com

Carl A. Nelson

Department of Mechanical
and Materials Engineering,
University of Nebraska–Lincoln,
Lincoln, NE 68508
e-mail: cnelson5@unl.edu

Manuscript received January 22, 2015; final manuscript received July 28, 2015; published online October 15, 2015. Assoc. Editor: Venketesh Dubey.

J. Med. Devices 9(4), 041012 (Oct 15, 2015) (14 pages) Paper No: MED-15-1015; doi: 10.1115/1.4031282 History: Received January 22, 2015; Revised July 28, 2015

Robotic minimally invasive surgery (R-MIS) has achieved success in various procedures; however, the lack of haptic feedback is considered by some to be a limiting factor. The typical method to acquire tool–tissue reaction forces is attaching force sensors on surgical tools, but this complicates sterilization and makes the tool bulky. This paper explores the feasibility of using motor current to estimate tool-tissue forces and demonstrates acceptable results in terms of time delay and accuracy. This sensorless force estimation method sheds new light on the possibility of equipping existing robotic surgical systems with haptic interfaces that require no sensors and are compatible with existing sterilization methods.

Copyright © 2015 by ASME
Your Session has timed out. Please sign back in to continue.

References

Figures

Grahic Jump Location
Fig. 1

A decoupled cable-driven grasper [18]

Grahic Jump Location
Fig. 2

Cables driving one jaw [18]

Grahic Jump Location
Fig. 3

Cable deformation model

Grahic Jump Location
Fig. 4

The 3DOF surgical grasper prototype

Grahic Jump Location
Fig. 6

Motion comparison between prototype and computer-aided design (CAD) model

Grahic Jump Location
Fig. 7

Force decoupling between grasp and yaw motion (* zero yaw angle is the position where the whole tool tip is in line with the tool shaft)

Grahic Jump Location
Fig. 9

Linear relation between clamp force and cable tension [19]

Grahic Jump Location
Fig. 10

Linear relation between motor current and motor torque

Grahic Jump Location
Fig. 11

The 3DOF surgical grasper prototype and master control

Grahic Jump Location
Fig. 12

Experiment setting of force estimation on (a) grasp DOF, (b) pitch DOF, and (c) yaw DOF [20]

Grahic Jump Location
Fig. 13

Force estimation on grasp DOF for long steady input

Grahic Jump Location
Fig. 14

Force estimation on grasp DOF for short steady input

Grahic Jump Location
Fig. 15

Force estimation on grasp DOF for periodic input [20]

Grahic Jump Location
Fig. 16

Influence of (a) pitch and (b) yaw motion on grasp force

Grahic Jump Location
Fig. 17

The actual sized prototype

Grahic Jump Location
Fig. 18

The force estimation fit before calibration on grasp DOF

Grahic Jump Location
Fig. 19

The calibrated result for long input on grasp DOF

Grahic Jump Location
Fig. 20

The calibrated result for short input on grasp DOF

Grahic Jump Location
Fig. 21

The calibrated result for periodic input on grasp DOF

Grahic Jump Location
Fig. 22

Stiffness differentiation

Grahic Jump Location
Fig. 23

Grasp force control (* every grid line on the x axis represents four seconds)

Grahic Jump Location
Fig. 24

Animal tissue with tumor embedded

Grahic Jump Location
Fig. 25

Grasp forces at locations (a) with tumor and (b) without tumor

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In