Preliminary Design and Engineering Evaluation of a Hydraulic Ankle–Foot Orthosis

[+] Author and Article Information
Brett Neubauer

Department of Mechanical Engineering,
University of Minnesota,
111 Church Street SE,
Minneapolis, MN 55455
e-mail: neub0070@umn.edu

William Durfee

Department of Mechanical Engineering,
University of Minnesota,
111 Church Street SE,
Minneapolis, MN 55455
e-mail: wkdurfee@umn.edu

Manuscript received June 23, 2015; final manuscript received March 22, 2016; published online August 5, 2016. Assoc. Editor: Rita M. Patterson.

J. Med. Devices 10(4), 041002 (Aug 05, 2016) (9 pages) Paper No: MED-15-1208; doi: 10.1115/1.4033327 History: Received June 23, 2015; Revised March 22, 2016

Ankle foot orthoses (AFOs) are used to correct motor impairments of the ankle. While current AFOs are passive, advances in technology and wearable robotics have opened the opportunity for a powered AFO. The hydraulic ankle foot orthosis (HAFO) is a device that takes advantage of the exceptional power-to-weight and force-to-weight of hydraulic fluid power. The device is untethered, and the power transmission chain is battery–electric motor–hydraulic pump–hose–cylinder, with the power supply worn at the waist and the cylinder actuators at the ankle. The fluid power circuit is configured as an electrohydraulic actuator (EHA) that is controlled by controlling the electric motor. The first prototype weighs 3.3 kg of which 0.97 kg is worn at the ankle. Steady-state torque–velocity performance showed that the prototype can provide 65 N·m of assistance torque and a no-load velocity of 105 deg/s. Closed-loop position control showed low steady-state error but a slow response. The current prototype demonstrates the potential of hydraulics for providing large torques in a compact, lightweight device. The speed performance of the prototype is inadequate for normal walking but can be improved by switching to servo valve control or by developing a custom hydraulic pump.

Copyright © 2016 by ASME
Your Session has timed out. Please sign back in to continue.

References

Uustal, H. , Baerga, E. , and Joki, J. , 2014, “ Prosthetics and Orthotics,” Physical Medicine and Rehabilitation Board Review, S. Cuccurullo , ed., Demos Medical Publishing, New York, Chap. 6.
National Institutes of Neurological Disorders and Stroke, 2009, “ NINDS Foot Drop Information,” National Institutes of Health, Bethesda, MD, accessed Dec. 28, 2015, http://www.ninds.nih.gov/disorders/foot_drop/foot_drop.htm
National Institutes of Neurological Disorders and Stroke, 2011, “ NINDS Spasticity Information,” National Institutes of Health, Bethesda, MD, accessed Dec. 28, 2015, http://www.ninds.nih.gov/disorders/spasticity/spasticity.htm
Lawler, B. , Oros, M. , and Wening, J. , 2007, “ A Functional Comparison of Solid and Articulated AFOs During Walking and Running in Children With Spastic Hemiplegic Cerebral Palsy,” ACPOC News, 13(2), pp. 5–18.
Ferris, D. P. , Czerniecki, J. M. , and Hannaford, B. , 2005, “ An Ankle-Foot Orthosis Powered by Artificial Pneumatic Muscles,” J. Appl. Biomech., 21(2), pp. 189–197. [CrossRef] [PubMed]
Shorter, K. A. , Kogler, G. F. , Loth, E. , Durfee, W. K. , and Hsiao-Wecksler, E. T. , 2011, “ A Portable Powered Ankle-Foot Orthosis for Rehabilitation,” J. Rehabil. Res. Dev., 48(4), pp. 459–472. [CrossRef] [PubMed]
Akers, A. , Gassman, M. , and Smith, R. , 2006, Hydraulic Power System Analysis, Taylor & Francis, Abingdon, UK.
Durfee, W. , and Sun, Z. , 2009, Fluid Power System Dynamics, Center for Compact and Efficient Fluid Power, Minneapolis, MN.
Hunt, T. M. , and Vaughan, N. , 1996, The Hydraulic Handbook, 9th ed., Elsevier, Philadelphia, PA.
Merritt, H. E. , 1967, Hydraulic Control Systems, Wiley, Hoboken, NJ.
Xia, J. , and Durfee, W. , 2013, “ Analysis of Small-Scale Hydraulic Actuation Systems,” ASME J. Mech. Des., 135(9), p. 091101. [CrossRef]
Manring, N. , 2005, Hydraulic Control Systems, Wiley, Hoboken, NJ.
Love, L. J. , Lind, R. F. , and Jansen, J. F. , 2009, “ Mesofluidic Actuation for Articulated Finger and Hand Prosthetics,” IEEE/RSJ Conference on Intelligent Robots and Systems (IROS), St. Louis, MO, Oct. 10–15, pp. 2586–2591.
Army Technology, 2014, “ Raytheon XOS 2 Exoskeleton, Second-Generation Robotics Suit,” Kable Intelligence Ltd., London, accessed Dec. 28, 2015, http://www.army-technology.com/projects/raytheon-xos-2-exoskeleton-us
Winter, D. A. , 2009, Biomechanics and Motor Control of Human Movement, Wiley, Hoboken, NJ.
Waters, R. L. , and Mulroy, S. , 1999, “ The Energy Expenditure of Normal and Pathologic Gait,” Gait Posture, 9(3), pp. 207–231. [CrossRef] [PubMed]
Skinner, H. , and Barrack, R. , 1990, “ Ankle Weighting Effect on Gait in Able-Bodied Adults,” Arch. Phys. Med. Rehabil., 71(2), pp. 112–115. [PubMed]
Tudor-Locke, C. E. , and Myers, A. M. , 2001, “ Methodological Considerations for Researchers and Practitioners Using Pedometers to Measure Physical (Ambulatory) Activity,” Res. Q. Exercise Sport, 72(1), pp. 1–12.
Tudor-Locke, C. , Johnson, W. D. , and Katzmarzyk, P. T. , 2009, “ Accelerometer-Determined Steps Per Day in U.S. Adults,” Med. Sci. Sports Exercise, 41(7), pp. 1384–1391. [CrossRef]
Tudor-Locke, C. , Craig, C. L. , Brown, W. J. , Clemes, S. A. , De Cocker, K. , Giles-Corti, B. , Hatano, Y. , Inoue, S. , Matsudo, S. M. , Mutrie, N. , Oppert, J. M. , Rowe, D. A. , Schmidt, M. D. , Schofield, G. M. , Spence, J. C. , Teixeira, P. J. , Tully, M. A. , and Blair, S. N. , 2011, “ How Many Steps/Day Are Enough? For Adults,” Int. J. Behav. Nutr. Phys. Act, 8, p. 79. [CrossRef] [PubMed]
Tarascon, J.-M. , and Armand, M. , 2001, “ Issues and Challenges Facing Rechargeable Lithium Batteries,” Nature, 414(6861), pp. 359–367. [CrossRef] [PubMed]
Love, L. J. , Richardson, B. , Lind, R. , Dehoff, R. , Peter, B. , Lowe, L. , and Blue, C. , 2013, “ Freeform Fluidics,” Mech. Eng., 135(6), p. S17.
Williams, L. , and Oak Ridge National Laboratory, 2013, “ ORNL-Group Shows How It’s Done, One Layer at a Time,” Additive Manufacturing, epub.
Lajeunesse, V. , Vincent, C. , Routhier, F. , Careau, E. , and Michaud, F. , 2015, “ Exoskeletons’ Design and Usefulness Evidence According to a Systematic Review of Lower Limb Exoskeletons Used for Functional Mobility by People With Spinal Cord Injury,” Disabil. Rehabil.: Assistive Technol., epub.
Hollander, K. W. , Ilg, R. , Sugar, T. G. , and Herring, D. , 2006, “ An Efficient Robotic Tendon for Gait Assistance,” ASME J. Biomech. Eng., 128(5), pp. 788–791. [CrossRef]
Mooney, L. M. , Rouse, E. J. , and Herr, H. M. , 2014, “ Autonomous Exoskeleton Reduces Metabolic Cost of Human Walking,” J. Neuroeng. Rehabil., 11(1), p. 151. [CrossRef] [PubMed]
Wiggin, M. B. , Sawicki, G. S. , and Collins, S. H. , 2011, “ An Exoskeleton Using Controlled Energy Storage and Release to Aid Ankle Propulsion,” IEEE International Conference on Rehabilitation Robotics (ICORR), Zurich, Switzerland, June 29–July 1.
Collins, S. H. , Wiggin, M. B. , and Sawicki, G. S. , 2015, “ Reducing the Energy Cost of Human Walking Using an Unpowered Exoskeleton,” Nature, 522(7555), pp. 212–215. [CrossRef] [PubMed]
Alexander, R. M. , 1988, Elastic Mechanisms in Animal Movement, Cambridge University Press, Cambridge, UK.
Fukunaga, T. , Kubo, K. , Kawakami, Y. , Fukashiro, S. , Kanehisa, H. , and Maganaris, C. N. , 2001, “ In Vivo Behaviour of Human Muscle Tendon During Walking,” Proc. R. Soc. London, Ser. B, 268(1464), pp. 229–233. [CrossRef]
Pedchenko, A. , and Barth, E. J. , 2009, “ Design and Validation of a High Energy Density Elastic Accumulator Using Polyurethane,” ASME Paper No. DSCC2009-2719.
Boes, M. K. , Islam, M. , Li, Y. D. , and Hsiao-Wecksler, E. T. , 2013, “ Fuel Efficiency of a Portable Powered Ankle-Foot Orthosis,” IEEE International Conference on Rehabilitation Robotics (ICORR), Seattle, WA, June 24–26.
Azizi, E. , Brainerd, E. L. , and Roberts, T. J. , 2008, “ Variable Gearing in Pennate Muscles,” Proc. Natl. Acad. Sci., 105(5), pp. 1745–1750. [CrossRef]
Azizi, E. , and Roberts, T. J. , 2013, “ Variable Gearing in a Biologically Inspired Pneumatic Actuator Array,” Bioinspiration Biomimetics, 8(2), p. 026002. [CrossRef] [PubMed]
Louie, D. R. , Eng, J. J. , and Lam, T. , 2015, “ Gait Speed Using Powered Robotic Exoskeletons After Spinal Cord Injury: A Systematic Review and Correlational Study,” J. Neuroeng. Rehabil., 12, p. 82. [CrossRef] [PubMed]
Forrester, L. W. , Roy, A. , Goodman, R. N. , Rietschel, J. , Barton, J. E. , Krebs, H. I. , and Macko, R. F. , 2012, “ Clinical Application of a Modular Ankle Robot for Stroke Rehabilitation,” NeuroRehabilitation, 33(1), pp. 85–97.
Mirelman, A. , Patritti, B. L. , Bonato, P. , and Deutsch, J. E. , 2010, “ Effects of Virtual Reality Training on Gait Biomechanics of Individuals Post-Stroke,” Gait Posture, 31(4), pp. 433–437. [CrossRef] [PubMed]
Zhang, M. , Davies, T. C. , and Xie, S. , 2013, “ Effectiveness of Robot-Assisted Therapy on Ankle Rehabilitation: A Systematic Review,” J. Neuroeng. Rehabil., 10(1), p. 30. [CrossRef] [PubMed]
Krebs, H. , Rossi, S. , Kim, S.-J. , Artemiadis, P. K. , Williams, D. , Castelli, E. , and Cappa, P. , 2011, “ Pediatric Anklebot,” IEEE International Conference on Rehabilitation Robotics (ICORR), Zurich, Switzerland, June 29–July 1.

Figures

Grahic Jump Location
Fig. 1

Fluid power circuit

Grahic Jump Location
Fig. 2

Hydraulic power supply

Grahic Jump Location
Fig. 4

Cross section of piston showing the cable rod

Grahic Jump Location
Fig. 5

Power supply prototype

Grahic Jump Location
Fig. 6

Ankle actuator (left) and power supply positioning on a user (right)

Grahic Jump Location
Fig. 7

Weight distribution for system components

Grahic Jump Location
Fig. 8

System and motor torque–velocity properties

Grahic Jump Location
Fig. 9

Open-loop unloaded velocity frequency response

Grahic Jump Location
Fig. 10

Open-loop isometric torque frequency response

Grahic Jump Location
Fig. 11

Open-loop isometric torque step response for system (solid) and motor (dashed) for torque steps of 20, 30, 40, and 50 N·m

Grahic Jump Location
Fig. 12

Closed-loop P control unloaded position step response

Grahic Jump Location
Fig. 13

Power supply mechanical efficiency excluding electric motor

Grahic Jump Location
Fig. 14

System mechanical (left) and volumetric (right) efficiencies excluding electric motor

Grahic Jump Location
Fig. 15

System total efficiency

Grahic Jump Location
Fig. 16

Torque–velocity of HAFO compared to torque–velocity of the ankle during one step

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In