Design and Evaluation of a Robotic Device for Automated Tail Vein Cannulations in Rodent Models

[+] Author and Article Information
Alex Fromholtz

Department of Biomedical Engineering,
Rutgers University,
Piscataway, NJ 08854
e-mail: afromholtz@gmail.com

Max L. Balter

Department of Biomedical Engineering,
Rutgers University,
Piscataway, NJ 08854
e-mail: balterm53@gmail.com

Alvin I. Chen

Department of Biomedical Engineering,
Rutgers University,
Piscataway, NJ 08854
e-mail: alv6688@gmail.com

Josh M. Leipheimer

Department of Biomedical Engineering,
Rutgers University,
Piscataway, NJ 08854
e-mail: joshleiph@gmail.com

Anil Shrirao

Department of Biomedical Engineering,
Rutgers University,
Piscataway, NJ 08854
e-mail: anilshrirao@gmail.com

Timothy J. Maguire

Department of Biomedical Engineering,
Rutgers University,
Piscataway, NJ 08854
e-mail: timjm@rutgers.edu

Martin L. Yarmush

Paul and Mary Monroe Distinguished Professor
Department of Biomedical Engineering,
Rutgers University,
Piscataway, NJ 08854
e-mail: yarmush@rci.rutgers.edu

1Corresponding author.

Manuscript received April 13, 2017; final manuscript received September 14, 2017; published online October 16, 2017. Assoc. Editor: Carl Nelson.

J. Med. Devices 11(4), 041008 (Oct 16, 2017) (7 pages) Paper No: MED-17-1204; doi: 10.1115/1.4038011 History: Received April 13, 2017; Revised September 14, 2017

Preclinical testing in rodent models is a ubiquitous part of modern biomedical research and commonly involves accessing the venous bloodstream for blood sampling and drug delivery. Manual tail vein cannulation is a time-consuming process and requires significant skill and training, particularly since improperly inserted needles can affect the experimental results and study outcomes. In this paper, we present a miniaturized, robotic medical device for automated, image-guided tail vein cannulations in rodent models. The device is composed of an actuated three degrees-of-freedom (DOFs) needle manipulator, three-dimensional (3D) near-infrared (NIR) stereo cameras, and an animal holding platform. Evaluating the system through a series of workspace simulations and free-space positioning tests, the device exhibited a sufficient work volume for the needle insertion task and submillimeter accuracy over the calibration targets. The results indicate that the device is capable of cannulating tail veins in rodent models as small as 0.3 mm in diameter, the smallest diameter vein required to target.

FIGURES IN THIS ARTICLE
<>
Copyright © 2017 by ASME
Your Session has timed out. Please sign back in to continue.

References

Nolan, P. M. , Peters, J. , Strivens, M. , Rogers, D. , Hagan, J. , Spurr, N. , Gray, I. C. , Vizor, L. , Brooker, D. , Whitehill, E. , Washbourne, R. , Hough, T. , Greenaway, S. , Hewitt, M. , Liu, X. , McCormack, S. , Pickford, K. , Selley, R. , Wells, C. , Tymowska-Lalanne, Z. , Roby, P. , Glenister, P. , Thornton, C. , Thaung, C. , Stevenson, J.-A. , Arkell, R. , Mburu, P. , Hardisty, R. , Kiernan, A. , Erven, A. , Steel, K. P. , Voegeling, S. , Guenet, J.-L. , Nickols, C. , Sadri, R. , Naase, M. , Isaacs, A. , Davies, K. , Browne, M. , Fisher, E. M. , Martin, J. , Rastan, S. , Brown, S. D. , and Hunter, J. , 2000, “ A Systematic, Genome-Wide, Phenotype-Driven Mutagenesis Programme for Gene Function Studies in the Mouse,” Nat. Genet., 25(4), pp. 440–443. [CrossRef] [PubMed]
Chang, Y.-C. , Berry-Pusey, B. , Yasin, R. , Vu, N. , Maraglia, B. , Chatziioannou, A. X. , and Tsao, T.-C. , 2015, “ An Automated Mouse Tail Vascular Access System by Vision and Pressure Feedback,” IEEE/ASME Trans. Mechatronics, 20(4), pp. 1616–1623. [CrossRef]
Springer, M. L. , Sievers, R. E. , Viswanathan, M. N. , Yee, M. S. , Foster, E. , Grossman, W. , and Yeghiazarians, Y. , 2005, “ Closed-Chest Cell Injections Into Mouse Myocardium Guided by High-Resolution Echocardiography,” Am. J. Physiol.: Heart Circ. Physiol., 289(3), pp. 1307–1314. [CrossRef]
Kanoh, K. , Shimura, T. , Suzuki, H. , Nomoto, K. , Asao, T. , and Kuwano, H. , 2004, “ Antitumor Effect of a Splenic Injection of 5-Fluorouracil on Metastatic Liver Cancer in Mice,” J. Pharmacol. Exp. Ther., 308(1), pp. 168–174. [CrossRef] [PubMed]
Malakoff, D. , 2000, “ The Rise of the Mouse, Biomedicine’s Model Mammal,” Science, 288(5464), pp. 248–253. [CrossRef] [PubMed]
Bebek, Ö. , Hwang, M. J. , and Cavusoglu, M. C. , 2013, “ Design of a Parallel Robot for Needle-Based Interventions on Small Animals,” IEEE/ASME Trans. Mechatronics, 18(1), pp. 62–73. [CrossRef]
Shima, H. , Ohno, K. , Michi, K-I. , Egawa, K. , and Takiguchi, R. , 1996, “ An Anatomical Study on the Forearm Vascular System,” J. Cranio-Maxillofac. Surg., 24(5), pp. 293–299. [CrossRef]
Staszyk, C. , Bohnet, W. , Gasse, H. , and Hackbarth, H. , 2003, “ Blood Vessels of the Rat Tail: A Histological Re-Examination With Respect to Blood Vessel Puncture Methods,” Lab. Anim., 37(2), pp. 121–125. [CrossRef] [PubMed]
Groman, E. V. , and Reinhardt, C. P. , 2004, “ Method to Quantify Tail Vein Injection Technique in Small Animals,” J. Am. Assoc. Lab. Anim. Sci., 43(1), pp. 35–38.
Gordon, C. J. , 1983, “ Influence of Heating Rate on Control of Heat Loss From the Tail in Mice,” Am. J. Physiol.: Regul. Integr. Comp. Physiol., 244(6), pp. R778–R784.
Berry-Pusey, B. N. , 2012, “ A Vascular Access System (VAS) for Preclinical Models,” Ph.D. thesis, University of California, Los Angeles, CA.
Singh, J. , 2012, “ The National Centre for the Replacement, Refinement, and Reduction of Animals in Research,” J. Pharmacol. Pharmacother., 3(1), pp. 87–89. [PubMed]
Waspe, A. C. , Lacefield, J. C. , and Fenster, A. , 2007, “ Registration of Three-Dimensional High-Frequency Ultrasound Images to a Robotic Needle-Positioning System for Pre-Clinical Research,” IEEE International Symposium on Biomedical Imaging: From Nano to Macro (ISBI), Arlington, VA, Apr. 12–15, pp. 1132–1135.
Berry-Pusey, B. N. , Chang, Y. C. , Prince, S. W. , Chu, K. , David, J. , Taschereau, R. , Silverman, R. W. , Williams, D. , Ladno, W. , Stout, D. , Tsao, T. C. , and Chatziioannou, A. , 2013, “ A Semi-Automated Vascular Access System for Preclinical Models,” Phys. Med. Biol., 58(16), pp. 5351–5362. [CrossRef] [PubMed]
Kazanzides, P. , Chang, J. , Iordachita, I. , Li, J. , Ling, C. , and Fichtinger, G. , 2007, “ Development of an Image-Guided Robot for Small Animal Research,” Comput. Aided Surg., 12(6), pp. 357–365. [CrossRef] [PubMed]
Waspe, A. C. , Cakiroglu, H. J. , Lacefield, J. C. , and Fenster, A. , 2007, “ Design, Calibration and Evaluation of a Robotic Needle-Positioning System for Small Animal Imaging Applications,” Phys. Med. Biol., 52(7), pp. 1863–1878. [CrossRef] [PubMed]
Ayadi, A. , Bayle, B. , Graebling, P. , and Gangloff, J. , 2008, “ An Image-Guided Robot for Needle Insertion in Small Animal. Accurate Needle Positioning Using Visual Servoing,” IEEE/RSJ International Conference on Intelligent Robots and Systems (IROS), Nice, France, Sept. 22–26, pp. 1453–1458.
Chen, A. I. , Balter, M. L. , Maguire, T. J. , and Yarmush, M. L. , 2016, “ 3D Near Infrared and Ultrasound Imaging of Peripheral Blood Vessels for Real-Time Localization and Needle Guidance,” International Conference on Medical Image Computing and Computer-Assisted Intervention (MICCAI), Quebec City, QC, Canada, Sept. 10–14, pp. 388–396.
Rand, R. , Burton, A. , and Ing, T. , 1965, “ The Tail of the Rat, in Temperature Regulation and Acclimatization,” Can. J. Physiol. Pharmacol., 43(2), pp. 257–267. [CrossRef] [PubMed]
Heikkila, J. , 2000, “ Using Circular Control Points,” IEEE Trans. Pattern Anal. Mach. Intell., 22(10), pp. 1066–1077. [CrossRef]
Balter, M. L. , Chen, A. I. , Maguire, T. J. , and Yarmush, M. L. , 2017, “ Adaptive Kinematic Control of a Robotic Venipuncture Device Based on Stereo Vision, Ultrasound, and Force Guidance,” IEEE Trans. Ind. Electron., 64(2), pp. 1626–1635. [CrossRef] [PubMed]

Figures

Grahic Jump Location
Fig. 1

Design of the robotic device for automated tail vein insertions in rodent models: (a) isometric view illustrating the needle manipulator assembly, imaging unit, rodent model, and animal holding platform and (b) side view of the device highlighting the imaging unit, LED array and motor controllers hidden under the holding platform

Grahic Jump Location
Fig. 2

Depiction of the three degrees of motion. (a) θFG and its relative coordinate frame on the holding platform. The axis of rotation is denoted as AoR. Additional components such as the light-emitting diode (LED) array and the tail clamp arm are also labeled. (b) Side and (c) front view of ZNM. This DOF is achieved using a rack and pinion mechanism with a helical pinion at 2 deg to align the motors with the insertion angle. (d) Dins and its chain of mechanisms linked together, which directly controls the needle insertion, and allows the needle path to remain parallel to the drive motor axle. Inset shows the slot design for finer control over mechanism backlash.

Grahic Jump Location
Fig. 3

Key components of the imaging system: (a) cross-sectional illustration of a rodent tail anatomy, (b) front (i) and side (ii) views of the imaging unit, with labeled camera angles, and (c) NIR tail vein imaging and segmentation. Image acquired under visible light (i), corresponding image acquired with the NIR imaging system using the transillumination method (ii), and tail vein segmentation applied to the NIR image (iii).

Grahic Jump Location
Fig. 4

Depiction of the actual (left) versus desired (right) workspace for the needle insertion task

Grahic Jump Location
Fig. 5

Free-space needle positioning tests: (a) the experimental setup using the cylindrical calibration rig and a secondary set of stereo cameras situated above the device as labeled in the figure, (b) stereo images from the left and right cameras during testing, and (c) images magnified 5×, and the needle tip manually selected to compute the 3D position

Grahic Jump Location
Fig. 6

Maps of mean absolute error interpolated over 25 calibration targets in (a) x, (b) y, (c) z, and (d) magnitude averaged across ten trials (units in millimeter). Origin corresponds to top left circle in calibration grid, and axes correspond to the global device coordinate frame.

Grahic Jump Location
Fig. 7

Low-level motor errors during position testing over 30 s interval, showing desired (solid line) versus actual (dotted line) signals for each DOF: (a) θFG, (b) ZNM, and (c) Dins

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In